Mass Flow Meter Calibration

At our calibration facility we carry out Mass Flow Meters (MFMs) in the range of 5,0 Sml / min - 50,0 Sl/min. The calibrations can be carried out with the following anhydrous gases: air and nitrogen. The flow rates indicated are expressed in reference to temperature and pressure standard conditions: 0 ° C and 101325 Pa.

The Mass Flow Meter (MFM) calibration takes place in our laboratory under controlled temperature and humidity conditions, in the following ranges:

  • Laboratory temperature: 23 °C ± 1 °C
  • Humidity: 30/70 %

The Mass Flow Meter (MFM) calibration is performed by comparing the sample flow measurement with our primary standard flow rate, according to the calibration method defined in the calibration procedures stored at our calibration laboratory.

We perform the calibration of Mass Flow Meter (MFM) from 3 to 10 calibration points between 5/10% and 100% of the measuring range of the device under test (DUT) with a minimum of 3 repetitions of the calibration test, up to 5 repetitions of the calibration test.

As result of the performed test, a calibration report will be issued containing:

  • Primary flow rate at standard reference conditions of 0 °C and 101325 Pa
  • Flow rate measured by the Mass Flow Meter under standard conditions of 0 ° C and 101325 Pa
  • Calibration coefficients and average calibration coefficient for calibration tests performed
  • Calibration uncertainty expressed as extended uncertainty of the calibration coefficients resulting from the tests performed
  • Environmental conditions during the calibration process
  • Reference to the procedures and the metrological chain used in the calibration process.
Facebook Twitter Google+ LinkedIn

Do you need a gas flow calibration? Contact us.

1000 characters left
Add files
When?

Monday - Friday
09:00 - 13:00


14:00 - 18:00

Where?

Via Padovane, 23/B
35010 - Massanzago (Padua)
ITALY

Contacts

tel. +39 (0)49 9811447
fax: +39 (0)49 7360352

Get in touch with us!

Linkedin
Twitter
Facebook

×

Log in